Total Organic Carbon
Description
3S Analyzers
Total Organic Carbon analyzer based on UV/persulfate method.
Method
The total organic carbon measurement is based on the UV/persulfate method. The sample is acidified and stripped with air to remove inorganic carbon. Sodium persulfate is added and the sample is processed in two UV reactors. In the reactors, the organic species present in the sample are completely oxidized to CO2. The resulting carbon dioxide is then measured by a NDIR detector and the TOC concentration is measured.
Key Differences
< click here
Total Organic Carbon analyzer based on UV/persulfate method.
Method
The total organic carbon measurement is based on the UV/persulfate method. The sample is acidified and stripped with air to remove inorganic carbon. Sodium persulfate is added and the sample is processed in two UV reactors. In the reactors the organic species present in the sample are completely oxidized to CO2. The resulting carbon dioxide is then measured by a NDIR detector and the TOC concentration is measured.
Technical Specification
< click here
Range
(with dilution) 20000 mg/l
Reproducibility ±2 % FS
Reproducibility (with dilution) ±4 % FS
Number of reagents 2
ModelTM
Sampling ModeContinuous.
SamplePressure: Atmospheric
Temperature: 2 - 70 °C (36 - 158 °F)
Flow Rate: 80 to 500 mL/min
Connection: 6 mm (¼-in.)
DrainPressure-free drain
Connection: 12 mm (½-in.)
Input Streams2
Cabinet MaterialEpoxy-coated stainless steel AISI 304
Dimensions (H x W x D)760 x 600 x 210 mm (29.6 x 23.4 x 8.2 in)
Weight37 kg (81 lbs)
Power SupplyInput Voltage:115 - 240 VAC 50/60 Hz
Power consumption: max. 350 W
Output signals2 x 4-20 mA analog outputs
Modbus RTU via RS485
Alarms2 SPDT programmable voltage-free relays
Digital InputOnline
Ambient Temperature5 - 45 °C (41 - 113 °F)
Ambient Humidity10 to 90 % RH (indoor use only, outdoor installation possible with protective cabinet or shelter not included)
Protection GradeIP54