The NIRS Analyzer Pro is a single point process analysis system based on high resolution diode array technology. It provides non-destructive analysis of liquids, granules, powders, slurries or opalescent substances by using fiber optics microbundles and contact probes for reflectance and transflectance measurements.

Key Differences

< click here

Technical Specification

< click here

NIRS Analyzer Pro
Analysis time: 5 - 50 ms / integration time; average time per result: 1 - 15 seconds.
Wavelength range: 1100 - 1650 nm
Detector: InGaAs diode spectrum
Spectral dispersion: 1.1 nm / pixel

General
Lifetime of the light source: Dual lamp system MTBF = 17,500 h
Software package (sold separately): Vision for instrument operation and method development
Wavelength accuracy: 0.5 nm
Wavelength precision: < 0.02 nm
Wavelength stability: < 0.01 nm/°C
Noise: < 60 Micro-AU
Random vibrations: 0.4 Grms at 10 - 150 Hz in accordance with IEC 60068-2-64 / 0.4 Grms at 10 - 1250 Hz in accordance with international Metrohm standard (additional information available on request)
Temperature: -5 - 40 °C (23 - 104 °F). With purge -5 - 65 °C (23 - 149 °F).
Purge: Flow rate min. 5 l/min, > 99.99% anhydrous, > 99.9% without oil and fine particles up to 0.3 μm
Humidity: 10 - 90% relative
Dimensions: (W × H × D) 42 × 42 × 13.5 cm (16.5 × 16.5 × 5.3 inches) + holders for the device
Weight: 25 kg / 55 lbs
Cabinet: 1.5 mm (lid 2.5 mm) stainless steel EN 1.4301 (SS2333)
Protection: IP69K 1) in accordance with IEC 60529 and DIN 40050 Part 9, NT ELEC 023
Power supply: disconnected from network or network-dependent (recommended) / 100 - 240 VAC, 50 - 60 Hz, 2.0 A, 150 W
1) IP6x is the highest protection against dust entering the device. IPx9K means protected against the effect of high-pressure water and/or steam cleaning at high temperature.