SPELEC VIS-NIR 350-1050 nm
Description
Metrohm Autolab
SPELEC is an instrument for performing spectroelectrochemical measurements. It combines in only one box a Lightsource, a Bipotentiostat/Galvanostat and a Spectrometer (UV-VIS wavelength range: 200-900 nm) and includes a dedicated spectroelectrochemical software that allows optical and electrochemical experiments synchronization.
Key Differences
< click here
Technical Specification
< click here
Current accuracy≤0.5% of current range at 100 nA to 10 mA
Current resolution0.1% of current output range
Dimensions in mm (W/H/D)24 x 11 x 25
Light Source (LS) wavelength range215-400 nm (deuterium) ; 360-2500 nm (tungsten halogen)
LS Bulb life>1,000 hours @ 240 nm (time); <50% @ 240 nm (decrease of intensity) ; Continuous operation (testing conditions)
LS Fiber optic connectorSMA 905
LS Ignition delay<2.0 seconds (delay for cold start-up may be longer)
LS Stability1.0% peak-to-peak (over 4 hours) after 30-minute warm-up
LS Time to stable output10 minutes (deuterium) ; 1 minute (tungsten halogen)
Maximum current in Ampère±40 mA
Maximum number of channels1
Maximum number of WE sharing AUX and REF2
Measured current resolution0.025 % of current range
Measured potential resolution0.012 % of potential range
Multichannel instrumentNO
Number of current ranges8
Number of current ranges remarks±1 nA to ±10 mA
Number of potential ranges (galvanostat)2
Number of potential ranges remarks (galvanostat)±100 mV to ±1 V
Operating modesBipotentiostat ; Potentiostat ; Galvanostat
PC interfaceUSB
Potential accuracy±0.2%
Potential range in volt±4V
Potential resolution1 mV
Power12 V DC
Spectrometer (SPM) DetectorLinear silicon CCD array
SPM A/D resolution16 bit
SPM Dark noise50 RMS counts
SPM Dynamic range8.5 x 10 rised to 7 (system) ; 1300:1 for a single acquisition
SPM Integration time1 ms to 65 seconds
SPM Optical resolution0.3+10.0 nm FWHM
SPM Pixel size14 µm x 200 µm
SPM Pixel well depth62,500 electrons
SPM Pixels2048
SPM Signal-to-noise ratio250:1 (at full signal)
SPM Stray light≤ 0.05 % at 600 nm ; < 0.10% at 435 nm
SPM Wavelength range200-900 nm
Weight of the instrument in kg1.95