SPELEC is an instrument for performing spectroelectrochemical measurements. It combines in only one box a Lightsource, a Bipotentiostat/Galvanostat and a Spectrometer (UV-VIS wavelength range: 200-900 nm) and includes a dedicated spectroelectrochemical software that allows optical and electrochemical experiments synchronization.

Key Differences

< click here

Technical Specification

< click here

Current accuracy≤0.5% of current range at 100 nA to 10 mA

Current resolution0.1% of current output range

Dimensions in mm (W/H/D)24 x 11 x 25

Light Source (LS) wavelength range215-400 nm (deuterium) ; 360-2500 nm (tungsten halogen)

LS Bulb life>1,000 hours @ 240 nm (time); <50% @ 240 nm (decrease of intensity) ; Continuous operation (testing conditions)

LS Fiber optic connectorSMA 905

LS Ignition delay<2.0 seconds (delay for cold start-up may be longer)

LS Stability1.0% peak-to-peak (over 4 hours) after 30-minute warm-up

LS Time to stable output10 minutes (deuterium) ; 1 minute (tungsten halogen)

Maximum current in Ampère±40 mA

Maximum number of channels1

Maximum number of WE sharing AUX and REF2

Measured current resolution0.025 % of current range

Measured potential resolution0.012 % of potential range

Multichannel instrumentNO

Number of current ranges8

Number of current ranges remarks±1 nA to ±10 mA

Number of potential ranges (galvanostat)2

Number of potential ranges remarks (galvanostat)±100 mV to ±1 V

Operating modesBipotentiostat ; Potentiostat ; Galvanostat

PC interfaceUSB

Potential accuracy±0.2%

Potential range in volt±4V

Potential resolution1 mV

Power12 V DC

Spectrometer (SPM) DetectorLinear silicon CCD array

SPM A/D resolution16 bit

SPM Dark noise50 RMS counts

SPM Dynamic range8.5 x 10 rised to 7 (system) ; 1300:1 for a single acquisition

SPM Integration time1 ms to 65 seconds

SPM Optical resolution0.3+10.0 nm FWHM

SPM Pixel size14 µm x 200 µm

SPM Pixel well depth62,500 electrons

SPM Pixels2048

SPM Signal-to-noise ratio250:1 (at full signal)

SPM Stray light≤ 0.05 % at 600 nm ; < 0.10% at 435 nm

SPM Wavelength range200-900 nm

Weight of the instrument in kg1.95