SPELEC UV-VIS 200-900 nm

BY
Metrohm Autolab
Metrohm Autolab
Description

SPELEC1050 is an instrument for performing spectroelectrochemical measurements. It combines in only one box a Lightsource, a Bipotentiostat/Galvanostat and a Spectrometer (VIS-NIR wavelength range: 350-1050 nm) and includes a dedicated spectroelectrochemical software that allows optical and electrochemical experiments synchronization.

Key Differences
Technical Specification

Current accuracy ≤0.5% of current range at 100 nA to 10 mA

Current resolution 0.1% of current output range

Dimensions in mm (W/H/D) 24 x 11 x 25

Light Source (LS) wavelength range 215-400 nm (deuterium) ; 360-2500 nm (tungsten halogen)

LS Bulb life >1,000 hours @ 240 nm (time) ; < 50% @ 240 nm (decrease of intensity) ; Continuous operation (testing conditions)

LS Fiber optic connector SMA 905

LS Ignition delay <2.0 seconds (delay for cold start-up may be longer)

LS Stability 1.0% peak-to-peak (over 4 hours) after 30-minute warm-up

LS Time to stable output 10 minutes (deuterium) ; 1 minute (tungsten halogen)

Maximum current in Ampère ±40 mA

Maximum number of channels 1

Maximum number of WE sharing AUX and REF 2

Measured current resolution 0.025 % of current range

Measured potential resolution 0.012 % of potential range

Multichannel instrument NO

Number of current ranges 8

Number of current ranges remarks ±1 nA to ±10 mA

Number of potential ranges (galvanostat) 2

Number of potential ranges remarks (galvanostat) ±100 mV to ±1 V

Operating modes Bipotentiostat ; Potentiostat ; Galvanostat

PC interface USB

Potential accuracy ±0.2%

Potential range in volt ±4V

Potential resolution 1 mV

Power 12 V DC

Spectrometer (SPM) Detector Linear silicon CCD array

SPM A/D resolution 16 bit

SPM Dark noise 50 RMS counts

SPM Dynamic range 8.5 x 10 rised to 7 (system) ; 1300:1 for a single acquisition

SPM Integration time 1 ms to 65 seconds

SPM Optical resolution 0.3+10.0 nm FWHM

SPM Pixel size 14 µm x 200 µm

SPM Pixel well depth 62,500 electrons

SPM Pixels 2048

SPM Signal-to-noise ratio 250:1 (at full signal)

SPM Stray light ≤0.05 % at 600 nm; <0.10 % at 435 nm

SPM Wavelength range 350-1050 nm

Weight of the instrument in kg 1.95